[Case 1. Digital refractor + LM]
1) Connect a digital refractor to COM1 of the RS232 Y CABLE.
2) Connect HLM to COM2 of the RS232 Y CABLE.
3) Set [BPS (COM1)], [RS232 PROTOCOL (COM1)] and [MODE (COM1)] for the target
digital refractor.
4) Set [LM PRINT] on.
5) Set [RS232 PROTOCOL (COM2)] on.
6) Set [BPS (COM2)] for the target HLM.
7) Select [LM] for [MODE (COM2)]
[Case 2. Digital refractor + LRK-Mate]
1) Connect a digital refractor to COM1 of the RS232 Y CABLE.
2) Connect LRK-Mate to COM2 of RS232 Y CABLE.
3) Set [BPS (COM1)], [RS232 PROTOCOL (COM1)] and [MODE (COM1)] for the target
digital refractor.
4) Set [LM PRINT] off.
5) Set [RS232 PROTOCOL (COM2)] on.
6) Set [BPS (COM2)] for the target LRK-Mate.
7) Select [Mate] for [MODE (COM2)]
[Case 3. Mate + LM]
1) Connect LRK-Mate to COM1 of the RS232 Y CABLE.
2) Connect LM to COM2 of the RS232 Y CABLE.
3) Set [BPS (COM1)], [RS232 PROTOCOL (COM1)] and [MODE (COM1)] for the target C
RK-Mate.