L98005G3-EU-A 15
User Manual: PCD-51™ 731A
Appendix II - EMC Manufacturer's Declarations
Electromagnetic Compatibility (EMC) statement for home healthcare
environment
The PCD-51™ system has been evaluated to international standard
IEC 60601-1-2 "General requirements for basic safety and essential
performance - Collateral standard: Electromagnetic compatibility -
Requirements and tests".
1. The Medical Electrical Equipment needs special precautions regarding
EMC and needs to be installed and put into service according to the
EMC information provided here within the accompanying documents.
2. Portable and mobile RF communication can affect the Medical
Electrical Equipment. See below recommended separation distances
between portable and mobile RF communication equipment and the
PCD-51™ system.
3. Wireless communication equipment such as wireless home network
devices, mobile phones, cordless telephones and their base stations,
walkie-talkies can affect this equipment and should be kept at least a
distance d = 3.3 m away from the system.
Rated Maximum Output
Power of Transmitter (W)
Separation Distance* according to
Frequency of Transmitter (m)
0.01 0.23
0.1 0.73
1 2.3
10 7.3
100 23
*Note: The distance calculated from 800 MHz to 2.5 GHz
Caution:
Do not use a mobile phone or other devices that emit
electromagnetic fields, near the unit. This may result in incorrect
operation of the unit.
Caution:
This system should not be used adjacent to or stacked with other
equipment. If adjacent or stacked use is necessary, observe the
device function to veri
fy normal operation in the configuration in
which it will be used