Model 731A - Electromagnetic Emissions - Manufacturer Declaration
The Model 731A is intended for use in the electromagnetic environment specified below. The customer or the user of the Model
731A should assure that it is used in such an environment.
Emissions Test Compliance Electromagnetic Environment - Guidance
RF emissions
CISPR 11
Group 1 The Model 731A uses RF energy only for its internal function. Therefore, its RF
emissions are very low and are not likely to cause any interference in nearby
electronic equipment.
RF emissions
CISPR 11
Class B
The Model 731A is suitable for use in all establishments, including domestic
establishments and those directly connected to the public low voltage power
supply network that supplies buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Complies
Voltage fluctuations/flicker
emissions
IEC 61000-3-3
Complies
Model 731A - Electromagnetic Immunity - Manufacturer Declaration
The Model 731A is intended for use in the electromagnetic environment specified below. The customer or the user of the Model
731A should assure that it is used in such an environment.
Immunity Test IEC 60601 Test Level Compliance Level Electromagnetic Environment - Guidance
Electrostatic discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
Complies
Electrical fast transient/burst
IEC 61000-4-4
±2 kV for power supply lines Complies Mains power quality should be that of a
typical domestic or hospital environment.
Surge
IEC 61000-4-5
±1 kV line to line Complies Mains power quality should be that of a
typical domestic or hospital environment.
Voltage dips, short
interruptions and voltage
variations on power supply
input lines
IEC 61000-4-11
>95% dip in U
T
for 0.5 cycle
60% dip in U
T
for 5 cycles
30% dip in U
T
for 25 cycles
>95% dip in U
T
for 5 cycles
Complies Mains power quality should be that of a
typical domestic or hospital environment.
Power frequency (50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m Complies Power frequency magnetic fields should be at
levels characteristic of a typical location in a
typical domestic l or hospital environment.
NOTE: U
T
is the AC mains voltage prior to application of the test level.