Status and error messages - The probe 38
Ringing! (during TEOAE
test only)
Poor probe fit. Reposition probe.
Stimulus too high (duing
ABR test only)
ear probe slipped during
ABR measure
Refit probe and restart test
Stimulus too high! Probefit not good.
•Probe not in ear.
Reposition probe.
System Error System malfunction. Contact your service depart-
ment.
Testing not possible! Impedance of electrodes or
balance is too high. EEG
values are constantly high.
Check electrodes and replace
them if necessary. Repeat test
if the patient is quiet.
Testing not possible!
Please disconnect the label
printer!
Testing cannot be done
while the printer is con-
nected to the device.
Disconnect the label printer
from the device.
Error message Cause Solution