Otometrics 37
Paper jam • Labels not correctly
adjusted.
•Adjust labels.
Poor testing conditions! • When there is too much
noise in more than 75% of
testing time, EEG and/or
interference is too high.
• Improve testing conditions.
• Reduce noise.
• Check interference from elec-
trical installations.
•Calm baby
Printer error • - • Check label printer manual.
Probe Error! • Calibration could not be
finished properly.
• Refit probe.
• Clean the probe tip.
• Check acoustic filter.
• Perform a probe test.
Error message Cause Solution