6.2 Additional Safety Specifications
6.2.1 Coil Temperature Protection
An indication of the coil temperature is given by the Coil Temperature information display on the User
Interface. If the coil surface temperature reaches its set limit, the display information will inform the
user that this condition has occurred. The normal operation of the equipment can then be resumed by
allowing the coil to cool, or alternatively, by replacing the coil with a cool one.
6.2.2 Other Safety Measures
If the instrument has not been triggered for over 10 minutes after the system status window has
displayed READY, the unit will automatically select the default mode and discharge internally The
Magstim Rapid
2
is capable of recognising automatically the type of stimulating coil being used.
6.3 Technical Specifications
Output Range and Accuracy:
0% - 100% of Maximum Voltage ±2%
Biphasic Single Cosine Cycle - 400µs period
Note: Magnetic field strength and maximum number of stimulations produced before coil exceeds
operating temperature are dependent on the stimulation coil being used. Please refer to the
appropriate coil operating manual.
Capacitor Life Expectancy:
2 x 10
6
discharges at 70% power level
8 x 10
5
discharges at 80% power level
4 x 10
5
discharges at 90% power level
2 x 10
5
discharges at 100% power level
All voltages are with respect to Aux Gnd signal on the isolated interface connecter
230V/240V Rapid² Systems: