Magstim Rapid²
Operating Manual
© The Magstim Company Limited 52 MOP03-EN-03
Calibration is not required.
6.4 Output Safety, Repetition Rate of Stimulus
The following Table is taken from Eric M. Wassermann’s paper – Risk and Safety of Repetitive
Transcranial Magnetic Stimulation: Report and recommendations from the International workshop on
the safety of Repetitive Transcranial Magnetic Stimulation June 5-7, 1996. It gives the maximum safe
duration (in seconds) of single trains of rTMS based on the NINDS experience.
Intensity (% of MEP threshold)
Numbers preceded by the symbol > are the longest durations tested. No after discharge or spread of
excitation has been encountered with single trains of rTMS at these combinations of stimulus
frequency and intensity.
6.5 Environmental Conditions*
Transport and storage temperature:
Operating, transport and storage relative humidity:
10% to 80% (non-condensing)
Operating atmospheric pressure:
Transport and storage atmospheric pressure: