Magstim Rapid²
Operating Manual
© The Magstim Company Limited 69 MOP03-EN-03
Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
The Magstim Rapid
2
System is intended for use in the electromagnetic environment specified below. The customer
or the user of the equipment should assure that it is used in such an environment.
Immunity Test
IEC 60601
Test Level
Compliance
Level
Electromagnetic environment - guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150 kHz to
80 MHz
3 V/m
80 MHz to
2,5 GHz
[V
1
] V
[E
1
] V/m
Portable and mobile RF communications
equipment should be used no closer to any part of
the Magstim Rapid
2
System, including cables, than
the recommended separation distance calculated
from the equation applicable to the frequency
transmitter.
Recommended separation distance
3,5
d = [ ] P
V
1
3,5 80 MHz to 800 MHz
d = [ ] P
E
1
7 800 MHz to 2,5 GHz
d = [ ] P
E
1
Where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended
separation distance in metres (m).
Field Strengths from fixed RF transmitters, as
determined by an electromagnetic site survey,
a
should be less than the compliance level in each
frequency range.
b
Interference may occur in the vicinity of equipment
marked with the following symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/ cordless) telephones and land
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with
accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey
should be considered. If the measured filed strength in the location in which the Magstim Rapid
2
System is used
exceeds the applicable RF compliance level above, the Magstim Rapid
2
System should be observed to verify
normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-
orienting or relocating the Magstim Rapid
2
System.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than [V
1
] V/m.