Operation Manual
touchTymp MI 26
and
MI 36
Version
4.2.7 The Built-In Printer
NOTE: This section only applies to touchTymp devices purchased with a built-in printer.
In order to change paper rolls:
• Push the marker on the left side of the touchTymp to open the printer cover (Figure 24).
• Pull the blue lever upwards (Figure 25).
• Insert a paper roll in the compartment with its loose end to the front of the printer and
the loose paper positioned underneath the roll as shown in the picture. Position the
loose end into the printer roll and hoist it by rotating the printer roll with your finger.
• Push the blue lever down. Close printer cover (Figure 26).
You can use the 0.5 ml, 1.0 ml, 2.0 ml and 5.0 ml test cavities
for validity check of the probe calibration (Figure 27). To
perform a probe check, select a protocol that measures a
tympanogram. Check the volume that was measured.
The allowed tolerance in the volume measurement is ±0.1 ml
for cavities up to 2.0 ml and ±5 % for larger cavities. These
tolerances are applicable for all probe tone frequencies.
NOTE: A probe check does not replace annual calibration
by your customer service. See also section 3.2.
4.2.9 Adjusting the Feet Height
Use the Allen key to adjust the touchTymp feet (Figure 28
and Figure 29).
NOTE: An Allen key is enclosed in the packaging of the eartip
box to enable adjustment of the pair of adjustable feet located
on the bottom of the touchTymp.
Please ensure that the Allen key is only used for the
purposes mentioned in this Operation Manual.