Operation Manual
touchTymp MI 26
and
MI 36
Version
Monitor phone: Allows audible presentation through the
Monitor phone of the device (Figure 150). These signals
include:
• Button touch: A click sound is delivered to the
monitor phone.
• Tone: The test signal is delivered to the monitor phone
as it is being delivered to the patient.
• Masking: Masking signal is delivered to the monitor
phone as it is being delivered to the patient.
5.6.16 Settings – Audiometry – Display
Display type: Selection of the test screen display (Figure
151). Choose between:
• None: The test screen removes any stored results as
no results can be stored within the device. Store
and NR buttons are removed from screen display
(Figure 152).
MI26 –None display MI 36 – None display
Figure 152
• Table: The test screen displays a table to store
numerical results. Results are plotted based on earside,
transducer and level at the time of selecting Store
/NR (Figure 153).
MI26 –Table display MI 36 – Table display
Figure 153