Bedpan / Urinal Bottle and Utensil / Bowl Washer Disinfector (WDS)
Operation, Maintenance and Installation Manual
Issue 16 Page 33 August 22
5.7 Making equipment safe after incomplete operating cycle
If an incomplete cycle has occurred, it is assumed that the load being processed has not been subjected to a complete
cycle as per the set process parameters. For this reason access to the load is restricted and will not be allowed until a full
cycle has been completed. At the end of an incomplete cycle or on powering up the device after an incomplete cycle, a
fault code will display relating to the cause of the incomplete cycle (Please see section 5.6 Fault Codes for description of
fault code). Depending on the nature of the fault a purge reset option may be available; this can be done by following the
instructions as per section 5.9 of this manual. If the device does not allow a purge reset to be completed, or the fault re-
occurs after a purge reset, it is recommended that the device is isolated and service is arranged to correct the fault.
5.8 In case of malfunction
In the event of a fault or malfunction with the device during operation, abort the cycle by pressing the Power On/Off
button on the control LCD. Once the device is off, isolate the mains power supply and both the hot & cold water services.
Notify maintenance or service provider of the fault and ensure device is not used until the fault is rectified.
5.9 Purge Reset (Operator)
1. Turn power OFF at control LCD display
2. Turn power ON at control LCD display
3. The display will alternate between the current fault/s and “PURGE”, press SELECT to start the Purge Reset cycle
When purge is complete the previous cycle will automatically start to ensure the load is disinfected, access to the load is
restricted until a cycle has been completed to the required parameters
If the fault reoccurs contact maintenance or service provider
5.10 Purge Detergent Line (Operator)
1. Turn power OFF at control LCD display
2. Turn power ON at control LCD display
3. The detergent line will automatically purge if necessary for a predetermined time or until the detergent is detected
If the detergent fault is not cleared check detergent level and repeat the purge process.
If the detergent continues to fault contact maintenance or service provider
The following functions must only be undertaken by Malmet service personnel or Malmet-trained facility maintenance
personnel.
5.11 Manual Purge Reset (Maintenance)
1. Turn power OFF at control LCD display
2. Turn power ON at control LCD display
3. Hold down both SCROLL and SELECT buttons, when MENU is displayed, release buttons
4. SCROLL to SERVICE menu and press SELECT
5. SCROLL to RESET menu and press SELECT CONFIRM’ should then display
6. Press SELECT again to start Purge Reset cycle
When purge reset is completed the display will resume normal operation