1600/1800 Manual
Press 【START/STOP】in the left one of Fig. 5 to go into testing
interface.(The right one of Fig 5)
Press 【START/STOP】again in the right one of Fig. 5, the testing
result displays on the screen, repeat this step, the results will
display on the screen one by one.
Print test report (For example: Print the report in “Basic mode”, the right one of Fig. 5)
Press
【
PRINT
】
in the right one of Fig.5 to print the report (Fig. 6)
Fig 6
Note: After printing, all the data displayed on the screen will disappear.
Before measurement
Make a blank reference solution by filling a clean cuvette (or test tube) half full with
distilled or de-ionized water or other specified solvent. Wipe the cuvette with tissue
to remove the fingerprints and droplets of liquid. Fit the blank cuvette into the 4-cell
linear changer and place the cuvette in the slot nearest you. For the MAPADA
UV/V-1600/1800, push the changer so that the cuvette is in the light path (Push the
rod in). Close the lid.
Analyze Sample
We provide three test methods in uv-16/18 series, they are basic mode test, quantitative
and kinetics.
1. Basic Mode
Push the blank cuvette into the light path. In main menu (the last one of Fig. 3), move the
cursor on “Basic mode”, then press
【
ENTER
】
to go into basic mode test. After
automatically blanking, it will display as the left one of Fig. 4 and wait for the operator.
【
RETURN
】
to exit.
Select Test Mode
There are three test modes (T%, Abs, Conc. / factor) for you to select by pressing
Basic Mode
NO. Abs. T%
1 0.000 100.0%
2 0.000 100.0%