1600/1800 Manual
11
2. Quantitative Mode
It includes two test methods, one is Standard curve, the other is Coefficient method. In the
former method, you can establish a standard curve using standard samples known
concentration. If you have known a curve equation, you can select the latter method.
Enter Quantitative Mode
Move the cursor on the Quantitative Mode, Press
【
ENTER
】
to go into the
method choosing interface.(Fig. 9)
ENTER
RETURN
Fig.9
2-1. Standard Curve Method
It’s a method to establish a standard curve by measuring a group of standard
samples.
Move the cursor on the Standard Curve, after 【
ENTER】 being expressed, it
goes into the pre-testing interface.(
((
(Fig.10)
))
)
ENTER
RETURN
Fig. 10
Set Wavelength
Press 【 GOTO λ 】 in the pre-testing interface, input the testing
wavelength by pressing the numeric keypad. (Fig.11). Reference Page
√
√√
√Quantitative
●
●●
●Standard Curve
○
○○
○Basic Mode
●
●●
●Quantitative
√
√√
√Quantitative:
::
:
●
●●
●Standard Curve
Standard Curve
0.000 Abs
0.000 ml/l
546.0 nm
Current WL.
Current Abs.
Concentration