EasyManua.ls Logo

MAPADA V-1600 - Page 24

MAPADA V-1600
36 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
1600/1800 Manual
23
Move the cursor on Lower Limit” by pressing
and
in setting
interface, followed byENTER(Fig.36)
ENTER
RETURN
Fig. 36
Input the new lower value by pressing the numeric keypad followed by
ENTER
, it will return to the kinetics setting interface.
If you don’t want to change the value being set last time, press ENTER
orRETURNdirectly .
Note: the range or A is -0.3-3, the range or T is 0-200, other value is invalid.
Sample Test
After you have finished setting “Time Interval”, “Total Test Time”, “Test
Mode”, “Upper Limit” and “Lower Limit”, press
RETURNto go back to
kinetics pre-testing interface. (The left interface of Fig. 37)
Pull the blank cuvette in the light path, press ZEROfor blanking.
Pull the sample cuvette in the light path, press STARTto go into testing
interface.
PressSTARTagain to began your test. (Fig. 37)
START
RETURN
Fig. 37
When the testing begins, the curve will be drawn on the screen in real
Kinetics
100.0 T%
546.0 nm
Kinetics 3.000
-0.500
0S 60S
546.0nm 0.002A
Kinetics
Upper Limit
Lower Limit
Kinetics: Abs.
Lower=-0.500
Please Input New
Lower=
DOMINIQUE DUTSCHER SAS