LF GENERATOR LEVEL
LFGL
LFGL?
:VALUE
Set LF Generator level (short form)
Set LF Generator
level
Datatype : Decimal
Numeric Program
Data
Allowed suffices : V,
MV, UV, DBMV
Default suffix : V
:INC Set LF Generator level step
Data type : Decimal Numeric Program Data
Allowed suffices : DB
Default suffix
:
DB
:UP Go UP one step
:DN Go DOWN one step
:RETN
Return
to original setting
:XFER Transfer current value to be the new setting
Data type
:
None
Allowed suffices
:
None
Default suffix : None
:UNITS
Select
default LF
level units
Data
type
:
Character
Program
Data
(DBM,
DBV, DBMV, V, MV,
or
UV)
Allowed suffices
:
None
Default suffix
:
None
Examples:
LFGL: VALUE 75.6MV;INC 20DB
Example:
LFGL: UP
Prepares the message containing information on LF Generator
Level setting
in the
following format:
:LFGL:UNITS<unit>;VALUE<nr2>;INC<nr2>
where: <unit> is character programmed data defining the
default LF level units (DBM,
DBV. DBMV.
V, MV
or UV).
: LFGL: UNITS MV; VALUE 125. 8; INC 1.0
3-2-30