49
Guidance and Manufacturer's Declarations - Electromagnetic Compliance
Electromagnetic Emissions
Electromagnetic Emissions
The ME Equipment is intended for use in the electromagnetic environment specified below. The customer
or the user of the ME Equipment should assure that it is used in such an environment.
Emission Test Compliance Electromagnetic Environment - Guidance
RF Emissions
(Radiated)
CISPR 11
Group 1
Class B
ME Equipment uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to cause
any interference in nearby electronic equipment.
RF Emissions
(Conducted)
CISPR 11
Group 1
Class B
Suitable for use in all establishments, including domestic
environments and those directly connected to the public low-voltage
power supply network that supplies buildings used for domestic
purposes.
Harmonic
Emissions
IEC 61000-3-2
Class A
Voltage
fluctuations/ Flicker
emissions
IEC 61000-3-3
Complies
Electromagnetic Immunity
Electromagnetic Immunity
The ME Equipment is intended for use in the electromagnetic environment specified below. The customer
or the user of the ME Equipment should assure that it is used in such an environment.
Immunity
Test
IEC 60601
Test Level
Compliance
Level
Electromagnetic Environment - Guidance
Electrostatic
discharge
(ESD)
IEC 61000-4-2
+/- 8 kV
contact
+/- 15 kV air
+/- 8 kV contact
+/- 15 air
Floors should be wood, concrete or ceramic tile. If
floors are covered with synthetic material, the relative
humidity should be at least 30%.