56
[3] The Masimo SET Technology has been validated for low perfusion accuracy in bench top testing
against a Biotek Index 2 simulator and Masimo’s simulator with signal strengths of greater than 0.02%
and transmission of greater than 5% for saturations ranging from 70% to 100%.
[4] The Masimo SET Technology has been validated for pulse rate accuracy for the range of 25-240 bpm
in bench top testing against a Biotek Index 2 simulator and Masimo’s simulator with signal strengths of
greater than 0.02% and transmission of greater than 5% for saturations ranging from 70% to 100%.
*Registered trademark of Fluke Biomedical Corporation, Everett, Washington.
Service and Maintenance
Cleaning
The Masimo Stork Hub is a reusable device. The devices are supplied and used non-sterile.
The Masimo Sensor should be removed from the Masimo Stork Boot and both cleaned separately before
and after they has been applied to a user and/or in accordance with local and governmental regulations
to minimize the risk of cross-contamination.
Smart Phone Cleaning
To properly clean the smart phone, refer to the smart phone's Operator's Manual or Directions For Use.
Masimo Stork Hub Cleaning
ALL TBD
CAUTION: Check the Home Medical Hub for possible cracks or opening before cleaning.
CAUTION: Do not allow liquids to enter the interior of the Masimo Stork Hub.
The outer surfaces can be cleaned either with a soft cloth dampened with a mild detergent and warm
water solution or they can be wiped down with the following cleaning solutions:
• 70% isopropyl alcohol
• 1:10 bleach to water solution (0.5% sodium hypochlorite)
• Super Sani-Cloth® Wipes (55% isopropyl alcohol, 0.5% quaternary ammonium chloride)
• Windex® (1.5% 2-Butoxyethanol, 1.5% ethylene glycol hexyl ether, 5% isopropyl alcohol)
• Formula 409® Antibacterial All-Purpose Cleaner (1.5% Lauramine oxide, 0.4% n-Alkyl
dimethyl benzyl ammonium chloride)
Masimo Sensor Cleaning
ALL TBD
WARNING: Before cleaning, make sure the sensor and chip are not applied to the hand.