McKesson TRUE METRIX® Control Solution Care
• Write date opened on control solution label. Discard
either 3 months aer rst opening or aer date
printed next to EXP on label has passed, whichever
comes rst.
• Aer use, wipe bottle tip clean and recap tightly.
• Store at temperature range of 40-86°F (4-30°C).
DO NOT FREEZE.
McKesson TRUE METRIX® PRO Professional
Monitoring Blood Glucose Test Strip Care
• Store test strips in original vial only. Do not transfer
old test strips to new vial or store test strips outside
of vial.
• Write date opened on test strip vial. Discard unused
test strips from vial either 4 months aer rst opening
or aer date printed next to EXP on label has passed.
Use of test strips past expiration dates may give
incorrect results.
• Close vial immediately aer removing test strip. Store
at temperature range of 40-86°F (4-30°C).
DO NOT FREEZE.
• Do not reuse test strip.
• Do not bend, cut or alter test strips in any way.