37
4. Let meter air dry thoroughly before
using to test.
5. Verify that the system is working
properly by performing an
Automatic Self-Test (see Getting
Started).
Stop using the meter if:
Meter display appears cloudy or any display segments
are missing,
Markings on meter, including back meter label, are
coming o or are missing,
Buttons are hard to push on the meter or do not work,
Unable to insert test strip into Test Port,
Automatic Self-Test gives an error message.
For general questions about McKesson TRUE METRIX®
PRO products, call 1-800-777-4908. For additional
assistance, call 1-800-803-6025.
Note: Other disinfectants have not been tested. e eect of
other disinfectants used interchangeably has not been
tested with the meter. Use of disinfectants other than
Super Sani Cloth Wipes may damage meter.
Note: Super Sani Cloth Wipes have been tested on the
meter for a total of 10,950 cleaning and disinfecting
cycles, which is equal to cleaning and disinfecting the
meter 10 times per day for a 3 year period. e use
life of the meter is 3 years.