Technical data
29
Electromagnetic immunity – for equipment and systems that are not life‑supporting
The FineTuner Echo is intended for use in the home healthcare environment. The
customer or the user of the FineTuner Echo should assure that it is used in such an
environment.
Immunity test IEC 60601 test level Compliance level
Electromagnetic environment –
guidance
Conducted RF
IEC 61000‑4‑6
3 Vrms
150 kHz to 80 MHz
Not applicable Portable and mobile RF communi‑
cations equipment should be used
no closer than 30 cm to any part of
the FineTuner Echo, including cables,
specied by MED‑EL.
Otherwise degradation of the
performance of the FineTuner Echo
could result.
6 Vrms in ISM and amateur
radio bands between
150 kHz and 80 MHz
80 % AM at 1 kHz
Radiated RF
IEC 61000‑4‑3
10 V/m
80 MHz to 2.7 GHz
10 V/m
Proximity elds
from RF wireless
communications
equipment
IEC 61000‑4‑3
27 V/m
380 MHz to 390 MHz
27 V/m
28 V/m
430 MHz to 470 MHz
28 V/m
9 V/m
704 MHz to 787 MHz
9 V/m
28 V/m
800 MHz to 960 MHz
28 V/m
28 V/m
1700 MHz to 1990 MHz
28 V/m
28 V/m
2400 MHz to 2570 MHz
28 V/m
9 V/m
5100 MHz to 5800 MHz
9 V/m