11
SYS02 board
2
J3
8
Socket to connect the
pressure&speed detection
board of the infusion pump
Connect J2 in the pressure
&speed detection board of the
infusion pump
3
J4
7
Socket to connect the
pressure&speed detection
board of the infusion pump
Connect J1 in the pressure
&speed detection board of the
infusion pump
4
J5
14
Socket to connect the
DC/DC board
Connect J2 in the DC/DC board
5
J7
4
Socket to connect the
pressure pressure&speed
detection board of the
infusion pump
Connect J7 in the
pressure&speed detection board
of the infusion pump
6
J8
14
Socket to connect the
DC/DC board
Connect J8 in the DC/DC board
7
J9
3
Not used
Apply in the infusion pump
8
J10
3
Socket to connect the
bubble sensor
Connect bubble sensor
9
J16
3
Socket to connect the
power supply
Connect the battery pack plug
10
CON1
8
Socket to connect the
interface board
Connect in the CON1 interface
board
2.4.4 Battery Pack
a. Introduction
SYS-6010 adopts the 11.1V/1500mA lithium-ion battery pack. Connect battery pack
to the power board with a connector. Disconnecting the connector, you can separate the
battery pack and connecting wire.
b. The Main Testing Points
No.
Testing
point
Description
Range
Unit
Remark
1
Black-blac
k wire
Short circuit test
The two wires
should be
shorted
2
Red-black
wire
Battery pack
voltage
9-12.5
V
Battery level influence
the voltage
Note: The quality of battery pack is not judged by its voltage. Connecting the infusion
pump to test the actual operating time is needed.
c. Number and Definition of the Sockets
No.
Socket
PIN
Description
Explanation