H4-EN Foreword P
AGE
10/44
1
1.6.3 Additional equipment
Other accessory devices, such as computers, may be interfaced to the Ergocard device. Using
accessory equipment that does not comply with the equivalent safety requirements of this
equipment may lead to a reduced level of safety and/or performance of the system.
Medisoft S.A. recommends that the user/customer employ the use of a line isolation
transformer with all accessory devices. Any accessories interfaced with this device must be
compliant with the IEC60601-1 and IEC 60601-1-2 standard.
For information about the connection of any additional equipment not listed below, please
contact Medisoft S.A. or its representative.
Printer
A printer can be used to print test results.
For ECG trace printing:
• Minimal required print speed: 15 ppm
• Recommended printer: Canon
Trolley
A trolley can be used to place the informatics equipment (computer, printer,...). The trolley
has to be medical grade and use an isolation transformer if the Ergocard is powered by the
trolley.
The positioning of the Ergocard module on a trolley not supplied by Medisoft S.A. is the
responsibility of the customer.
Computer
A computer is intended to be used with the device in order to run the Expair software. The
choice of the computer used is the responsibility of the customer.
Minimal required specifications:
• Operating system: Windows 7 Pro 32/64 bits, Windows 8 Pro (8.1) 32/64 bits,
Windows 10
• CPU: minimum 3.5 GHz
• Hard disk: minimum 500 Gb
• RAM: DDR, minimum 4 Gb, minimum 1666 MHz
• 1 SVGA graphics port or HDMI
• 4 USB 2.0 ports minimum
• 2 USB 3.0 ports minimum
• Keyboard / mouse / touch screen
Heart rate instrument
A heart rate instrument can be used to measure heart rate.
Approved device: Zephyr HXM BT (wireless Bluetooth heart rate monitor)