H4-EN Calibration P
AGE
34/44
5
On the calibration menu, select the Ergo-CrfN
2
tab.
If two gas cylinders are used (a calibration gas cylinder and a zero gas cylinder),
make sure the check box Enable ambient gas cylinder is ticked.
Click on the button Certificate 1 in order to access the calibration gas cylinder
details.
Check and edit the fields if the calibration gas cylinder has been changed. All
information is available on the analysis certificate that comes with the cylinder.
Click on the button Certificate 2 in order to access the ambient gas cylinder details
(only if this cylinder is enabled).
Check and edit the fields if the ambient gas cylinder has been changed.
Click on Calibrate. Wait for the system to complete the calibration.
Select More. Check the results of the calibration.