15. Technical specifications
Test voltage : Up to 80,000 V.
Test voltage resolution : 100 V between 10 kV and 80 kV.
10V up to 9.99 kV.
Test voltage accuracy : ± (2% of nominal value ± 2 digits) with 1 mA
load.
Polarity : Negative voltage, ground positive.
Leakage current : Max. 10 mA.
Leakage current resolution : 0.01 mA.
Leakage current accuracy : ± (2% of nominal value ± 2 digits).
Display : Alphanumeric.
Ripple : < 2%.
Safety features : Remote control via Bluetooth.
Emergency stop buttons.
Self-detection of sparks.
Visual indicators (LEDs).
Sound indicator.
PC Interface
: USB.
Advanced features
: Ramp and SVT test.
“Pass-fail” and fixed time tests
Internal memory.
Built-in chronometer
:
Indicates elapsed time from the beginning of
the measurement mm:ss format, up to 20:00.
Software MegaLogg2 : It allows to transfer the data stored in the
equipment memory to a PC, to analyze
them, to present the results in tables and
graphs, to generate reports of the tests.
Environmental protection
index
: IP54 (with the cover closed).
Safety
: Conforms to IEC 61010-1.