EasyManua.ls Logo

Megger DLRO10 - Data, Battery, and Environmental Specifications; Safety, EMC, Dimensions, and Weight

Megger DLRO10
26 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
www.megger.com
DLRO10 and DLRO10X
20
Data Transfer Real Time or from storage via RS232
Storage 700 tests
Memo Field Up to 200 characters per test via integral
alphanumeric keypad.
Battery Capacity 5.2 Ah Li ion
Life Typically 1300 x 10 A tests before recharge
Recharge 12 to 15 V d.c. supply
Charging Time 4 hrs for full charge
Temperature Operation +5 °C to +45 °C (41 °F to 113 °F) at full specification.
-10 °C to +50 °C (14 °F to 122 °F) at reduced accuracy
Storage -30 °C to +70 °C (-22 °F to 158 °F)
Calibration 20 °C (68 °F)
Coefficient <0.01% per °C from 5 °C to 40 °C (<0.006% per °F from 41 °F to 104 °F)
Standard charging 0 °C to +40 °C (32 °F to 104 °F)
Humidity Max 90% RH @ 40 °C (104 °F) non-condensing
Altitude Max 2000 m (6562 ft.) to full safety specifications
Safety In accordance with IEC61010-1 600 V Category III - only when DH6 leads are used
EMC In accordance with IEC61326-1
Dimensions (HxWxD) With Battery
245 x 237 x 100 mm (9.6 x 9.3 x 4 inch)
Without Battery
171 x 237 x 100 mm (6.7 x 9.3 x 4 inch)
Weight With Battery 2.46 kg (5.42 lb)
Without Battery
1.48 kg (3.26 lb)

Related product manuals