Megger MIT417 - Voltage Range; Power Supply; Battery Life

Megger MIT417
14 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Leakage current: 10% ±3 digits
Voltage display: 3% ±3 digits ±0.5% of rated
voltage
Dielectric absorption Ratio (DAR): 60sec / 30sec ratio
Notes:
(1) All ranges measure from 0,00 MΩ upwards.
(2) Above specifications only apply when high quality
silicone leads are being used.
Continuity
EN61557 operating range: 0,01 Ω to 99,9 Ω (0 to 100 Ω on
analogue scale)
Accuracy: ± 2% ± 2 digits (0 to 100 Ω)
Open circuit voltage: 5 V ± 1 V
Test current: 205 mA (±5 mA) (0.01 Ω to 9.99 Ω)
20 mA (±1 mA) (10.0 Ω to 99.9 Ω)
Lead resistance zeroing: Up to 9,99 Ω
Buzzer:
Variable limit 1 Ω, 2 Ω, 5 Ω, 10 Ω, 20 Ω
Resistance
EN61557 operating range:
0,01 kΩ to 1000 kΩ (0 to 1 MΩ on analogue scale)
Accuracy: ±3% up to 50 kΩ then ±5% ±2 digits
Open circuit voltage: 5 V ±1 V
Short circuit current: 1.5 mA ±0.2 mA
Voltage range
0 to 600 V d.c. ± 2% ± 2 digits
10 mV to 600 V TRMS sinusoidal (40 – 400 Hz) ±2% ±2 digits
0 to 1000 V on analogue scale
Unspecified input level 0 – 10 mV (40 – 400 Hz)
For non sinusoidal waveforms additional specifications apply:
±3% ±2 digits 101 mV – 600 V TRMS and
±8% ±2 digits 10 mV – 100 mV TRMS
Default Voltmeter
Operates at >25 volts a.c. or d.c., on any range except OFF
Frequency: 15-400 Hz (15 Hz - 99,9 Hz) ±0.5% ±1 digit
(100 Hz to 400 Hz)
Power supply
5 x 1,5 V cells type IEC LR6 (AA, MN1500, HP7, AM3 R6HP) Alkaline
NiMH rechargeable cells may be used.
Battery life
2200 insulation tests with duty cycle of 5 sec on 55 sec off