EasyManuals Logo

Megger MIT481/2 User Manual

Megger MIT481/2
45 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #42 background imageLoading...
Page #42 background image
www.megger.comMIT400/2
42
15.3 Blown fuse indicator
The blown fuse indicator is shown below:
z
f
100 0
8 00
6 004 00
2 00
0
W
W
Am
- Figure 48
This symbol operates on the continuity [W] range. The symbol indicates that one of the fuses in the instrument has failed.
15.4 Fuse replacement procedure
1. Switch off the instrument and disconnect the instrument from any electrical circuits.
2. Disconnect all test leads from the instrument.
3. Remove the rear fuse cover. To remove the rear cover, undo the screw on the rear of the small fuse cover and lift the cover off,
as below:
- Figure 49
4. Both fuses should be checked for failure and replaced if faulty.
A replacement fuse must be of the correct type and rating: i.e. 500 mA (FF) H.B.C.30 kA min 1000 V (32 mm x 6 mm).
5. Replace the cover and retaining screw.
15.5 Preventive maintenance
The MIT400/2 series instruments require very little maintenance.
Test leads should be checked before use to ensure there is no damage.
Ensure batteries are removed if the instrument is left unused for an extended period.
When necessary, the instrument can be cleaned with a damp cloth.
Do not use alcohol based cleaners as these may leave a residue.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Megger MIT481/2 and is the answer not in the manual?

Megger MIT481/2 Specifications

General IconGeneral
BrandMegger
ModelMIT481/2
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals