11
Safety: Meets the safety requirements for double insulation to IEC 1010-1 (1995) EN61010 (1995) to
installation Category III***, 300 Volts phase to earth (ground) or 600 Volts Category I*
Non replaceable Fuse: 1 Amp, 250 V, HBC type (F) (20 mm x 5 mm) to IEC 127/1
This fuse protects the instrument against any faults occurring when using rechargable batteries.
E.M.C: Meets EN 50081-1 and EN 50082-1 (1992) with a maximum error of ± 5% f.s.d. in areas of high
R . F.
Temperature Range:
Operating: 0˚C to 30˚C (32˚F to 86˚F) at full specification
Operating: -20˚C
†
to 50˚C (-4˚F to 122˚F) to temperature coefficient ±0,1% / ˚C
Storage: -25˚C to 65˚C (-13˚F to 149˚F)
Humidity: 90% RH maximum at 40˚C (104˚F)
Dimensions: 220 mm x 160 mm x 115 mm (8,9 in x 6,3 in x 4,5 in)
Weight: BM15 Approx 1.2 kg
MJ15 Approx 1,6 kg, or 1,8 kg with battery holder and cells.
Cleaning: Wipe the disconnected instrument with a clean cloth dampened with soapy water or Isopropyl
Alcohol (IPA)
***Relates to transient overvoltage likely to be found in fixed installation wiring.
*Relates to transient overvoltage likely to be found in special equipment or parts of equipment, telecommunication, electronic etc.
† Note: MJ15 will operate as normal by generator at full temperature range
BM15 will require new healthy battery cells to operate as normal at -20˚C
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com