Fuses: 500 mA (FF) 660 V Ceramic 50 kA HBC 1
1
/4 ins x
1
/4 ins (32 mm x 6 mm)
7A (F) 440 V Ceramic 10kA HBC 1
1
4 ins x
1
4 ins (32 mm x 6 mm)
100 mA (F) HBC 20 mm x 5 mm (for line protection only)
MJ359 and MJ559 only: Power connection plug fuse - 100 mA 240 V HBC
25
32 ins x
1
4 ins (20 mm x 6 mm)
Mains power cord fused plug (when applicable) 3 A 250 V ceramic H.B.C. fuse to
BS1362 1
1
4 ins x
1
4 ins (32 mm x 6 mm)
E.M.C. The instruments meet EN 50081-1 and EN 50082-1 (1992)
Operating Temp. Range: 32˚ to 113˚F (0˚ to 45˚C) for battery operated models
14˚ to 122˚F (-10˚ to 50˚C)
Humidity Range
Operating: 70% RH max. at 68˚F, (20˚C)
60% RH max. at 95˚F, (35˚C)
50% RH max. at 105˚F (40˚C)
Storage: 95% R.H. max. at 95˚F (35˚C)
Dimensions
Length: 180 mm (7 ins) (210 mm [8,3 ins] including generator handle)
Width: 125 mm (4,9 ins)
Height: 130 mm (5,1 ins)
Weight: Approximately 1 kg (2,3lb)
Cleaning: Wipe disconnected instrument with a clean cloth dampened with soapy water or