18
Technical Specification
OTS100AF/80AF/60AF
The following technical specifications apply to the OTS100AF/80AF/60AF:
Test voltage -50 to 0 to 50 kV, -40 to 0 to 40 kV, -30 to 0 to 30 kV @ 61.8 Hz
Voltage resolution 0.1 kV
Voltage accuracy ±1% ±2 digits
Vessels
400 ml (standard)
150 ml (option) Thermoplastic polyamide chamber provides precision electrode
alignment, adjustment wheels lock electrodes in position
Temperature sensor resolution 1 ºC
Power supply Line voltage/power 90 to 264 VAC, 300 VA
Line frequency 48 / 63 Hz
Fuses 2 x 4 A CFST ceramic, 20 mm x 5 mm, for 250 V
Internal printer (option) Dot matrix impact printer with graphical output
Paper 57.5 mm wide, weight 52 gm per roll
Epson black ribbon cassette ERC-09B
Protection Safety interlock on test chamber
Discharge barrier on polycarbonate test chamber cover
Display 320 x 240 QVGA colour display with backlight
Dimensions (W x H x D) OTS100AF/80AF/60AF 520 mm x 420 mm x 290 mm
Weight OTS100AF/80AF/60AF 30 kg
Operating temperature 0 ºC to +50 ºC
Storage temperature -30 ºC to +65 ºC
Humidity 80% RH at 40 ºC operation
95% RH at 40 ºC storage
Altitude 1000 m
Safety Designed in accordance with IEC61010
EMC Light industrial IEC 61326-1 Class B, CISPR 22, CISPR 16-1 and CISPR
16-2