4.7 Pump internal flow
sanitization
Material needed:
1. Sanitization kit (item MMSANKIT1)
2. Consumables for sanitization kit
(item MMSANSYFU)
3. Sanitizer (see table)
4. Purified water
• Remove the pump heads
• Place a sanitization kit on one pump head
support and lock it
• Lock the sanitization kit plug on the
other support
Optional: For faster sanitization, it is
possible to replace the plug by a second
sanitization kit and sanitize the complete
pump at once.
• Rotate the trigger counter clockwise,
the switch turns green
Figure 20: Trigger is rotated counter clockwise
• Take one sanitization funnel assembly
(sanitization kit consumable) and
remove its support, then place the
funnel on the kit
• Take a syringe (sanitization kit
consumable) and remove its plunger
• Place and lock the syringe on the kit
• Pour 60 mL of sanitizer in the syringe
• Introduce the plunger into the syringe
Note: An air pocket is visible over the
liquid.
Figure 21: Visible air pocket over the liquid in the
sanitization syringe
• Push the plunger to bring the liquid level
to the 50 mL level indicator
Figure 22: Plunger pushed to bring the liquid
to the 50 mL level indication of the sanitization
syringe
• Open the funnel lid
• Pour puried water up to the 250 mL
indication level
• Put back the funnel lid
• Touch the switch to start the pump
Note: the switch starts blinking for
15 minutes while the following
operations occur:
1. The plunger is partially pulled down
and the sanitizer fills the pump internal
flow.
2. About 10 mL of sanitizer remains into
the syringe
3. The sanitizer remains in contact with
the internal flow for 15 minutes