EasyManua.ls Logo

Merck Milliflex Oasis - Using Multiple Milliflex Oasis Pumps; Material Needed for Multiple Pump Setup; Installation for Multiple Pump Systems

Merck Milliflex Oasis
24 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Page 18 of 24
6. Use of multiple pumps
Up to three pumps can be connected
providing the following benefits:
• Higher throughput
• Use of single power supply
• Daisy chain start: up to six samples
filtered with a single push of a button
6.1 Material needed
6.1.1 Two pumps connected:
• Two Milliflex Oasis
®
pumps for EZ-Fit
®
filtration unit (item MMSYSTEZ1)
• Milliflex Oasis
®
electrical cable
for connecting two pumps
(item MMCABLEMM)
• Milliflex Oasis
®
power supply, the item
number is specific to defined regions:
- Australia: item MMPWRSPAU
- Brazil: item MMPWRSPBR
- China: item MMPWRSPCN
- Denmark: item MMPWRSPDK
- Europe: item MMPWRSPEU
- Switzerland: item MMPWRSPSZ
- India: item MMPWRSPIN
- UK: item MMPWRSPUK
- Japan: item MMPWRSPJP
- South Africa: item MMPWRSPZA
Optional: Milliflex Oasis
®
T connectors for
tubing (item MMTCNNECT)
6.1.2 Three pumps connected:
• Three Milliflex Oasis
®
pumps for EZ-Fit
®
filtration unit (item MMSYSTEZ1)
• Two Milliflex Oasis
®
electrical cables
for connecting two pumps (item
MMCABLEMM)
• Milliflex Oasis
®
power supply, the item
number is specific to defined regions
Optional: Milliflex Oasis
®
T connectors for
tubing (item MMTCNNECT)
6.2 Installation
The pumps must be installed on a level
surface to ensure good sample volume
readings.
• Take the pump bodies and the tubing
out of the package with care
• Take out the accessory bags
Optional: the accessory bag contains four
stickers, select your preferred sticker and
stick it carefully to the specific groove on
each pump.
Figure 30: Pump sticker set up
• Clean the exterior of the pumps and
the tubing using a wipe moistened with
a surface sanitizer listed in the cleaning
& maintenance section
• Place the pumps in the working area
Under a horizontal laminar flow hood, the
front of the pumps must be placed at least
at 160 mm (6.3 in.) from the front edge of
the LFH (see picture).
Figure 31: The front of the pump must be placed
at least at 160 mm (6.3 in.) from the front edge
of a horizontal LFH

Related product manuals