4.10 Sanitization kit
gaskets (spare parts)
replacement
Materials:
1. Sanitization kit (item MMSANKIT1)
2. Sanitization kit gasket kit (item
MMGASKTSK) containing:
• Five O-rings for securing funnel to
kit tightness
• Five O-rings for securing kit to pump
tightness
• Remove the gaskets (using the metal
tool provided with the gasket kit) from
the sanitization kit and discard them
• Take out a set of gaskets out of the
gasket kit bag and clean them using a
wipe moistened with a surface sanitizer
listed in the cleaning & maintenance
section
• Put the new gaskets in place