METEK GmbH uSonic-2
24
Example:
AV=4, AT=10, SF=40000
4 measuring cycles (40Hz) will be averaged coherent and committed to the
processor. The processor averages these results and displays them after 10s.
The processor can handle max. 10 Hz. Consequently the commands SF und AV
are depending of each other and may limit each other.
The formula for valid settings is:
SF/AV=max. 10000 and SF=max 40000.
Invalid settings (for ex. AV=1 and SF=40000) are blocked and the message
„parameter out of range“ will appear.