METEK GmbH uSonic-2
AE=3 C1…C4 maximum value
See chapter 4.2.4
AO Analog Offset
Possible setting 0…50 (% of the total available range)
The command AO configures an offset to the analog output signals.
This offset configures a range, which is not used for the data output.
The not used range can be used for an error indication.
See chapter 4.2.1.
AT Averaging Time
Possible setting 0 ... 3600 (s)
The parameter AT defines the averaging time.
See chapter 3.3.4
AV Averaging Number
Possible setting 1 ... SF/1000, max. 32
The command AV defines, if and how many measured values will
averaged coherent. The coherent averaging function averages the
measured signals directly, before sending these results to the
processor. The coherent averaging is recommended for high wind
speed application.
See chapter 3.3.4
AZ Device Azimuth
Possible setting 0 ... 359