METEK GmbH uSonic-2 45
data output @ 11:07:15h
data output @ 11:07:30h etc.
If SY has been set to 1 while the residual interval
period is less than 50 % of the nominal interval the
interval is extended to the next complete interval.
TC Calibration Temperature
Possible setting -5000 ... 5000 (0,01°C)
To calibrate the uSonic-2, it is necessary to measure the temperature
with a thermometer and enter the temperature with the command TC.
The calibration temperature can just be entered in the service mode.
Example:
TC=2150 : Calibration temperature is 21,5 °C
See appendix B, Calibration procedure.
TI Time and Date
The command TI sets the date and time of the internal clock of the
uSonic-2. The internal clock is battery- buffered.
Format: dd.mm.yy hh:mm:ss
Example:
TI=13.10.11 14:07:00 ENTER
The date and time is set to 13.Oct.2011, 14:07:00
The date and time can just be entered in the service mode.
TV Test Values
Possible setting 0 ... 100 (%) and 101
The analog output ports can be checked with the test values. If
OA=32, the analog output signals are defined with a fixed value.