METEK GmbH uSonic-2
on automatically. This may also happen if the sensor is started while stored in a
box with reflections which causes implausible data! HT=2 should be used
exclusively after the sensor has been installed at site.
LC Last Calibration
This information displays the date and time of the last calibration and will be set
by the uSonic-2 while calibration procedure and cannot be entered by the user.
LD Log Data
Possible setting 0, 1, 2, 55 or 99
The command LD defines the status of the internal data logger
LD=0 stop logging data
LD=1 start logging data, no output of data on the serial line
LD=2 start logging data and output of data on the serial line
LD=55 set back marker for unread data and start logging data
LD=99 erase of internal data logger and start logging data
See chapter 7
LF Logger free
Shows the number of free data sets of the internal data logger.
LT Logger total
Shows the total number of data sets of the internal data logger.