METEK GmbH uSonic-2 49
The stored configurations can be recalled with the commands L1…L3.
The actual configuration is automatically stored in the internal
See also description L1…L3
SV Service mode
The command SV starts the service mode for 1 minute.
The started service mode is required to enter the following commands:
P1…P3 Enter the path lengths
TC Enter the calibration temperature
CA Start the calibration
TI Set date and time
storage #0. This storage won’t be affected by the L1… and S1…
commands.
Example:
The commands N0=20110928 and S1 will store the actual configuration
with the name 20110928. If the uSonic-2s configuration is changed and
the user wants to change back to the stored configuration, the user just
needs to enter L1 and the uSonic-2 works in the stored configuration.
VS Version number
The command VS reads out the several system parameters.
Example:
C:VS
R:SN serial number of the sonic
R:VN number of the firmware version
R:VD date of the generation of the firmware version