METEK GmbH uSonic-2
Examples:
TV=0 uSonic-2 variant 0...10 VDC / 0...20 mA
AO=0 C1, C2 value: 0 VDC
OA=32 C3, C4 value: 0 mA
TV=50 uSonic-2 variant 0...5 VDC / 0...20 mA
AO=0 C1, C2 value: 2.5 VDC
OA=32 C3, C4 value: 10 mA
The additional function TV=101, defines the analog output ports C1...C4 with
different values to check the alignment.
Example:
TV=101 uSonic-2 variant 0...10 VDC / 0...20 mA
AO=0 Port1 value 2.5 VDC (25%)
OA=32 Port2 value 5.0 VDC (50%)
Port3 value 15 mA (75%)
Port4 value 20 mA (100%)
Attend also the command OA (see chapter 4.2.1) in this case.