EasyManua.ls Logo

Metek uSonic-2 - Possible Settings for Command NO

Default Icon
58 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
METEK GmbH uSonic-2 31
The output of this sentence can be activated/deactivated according to the setting
of parameter NO (refer to table in chapter 5.1). Additional a checksum can be
applied to each sentence (MWV and MTA). This behaviour is selected by
parameter FR. If this parameter is set to 1 the checksum is appended to the end
of each sentence by use of the delimiter "*" and appears after this character as a
two digit hexadecimal value as defined in the NMEA0183 standard. The value is
calculated by XOR'ing the 8 binary data bits of each character of the sentence
between the leading "$" and the delimiting "*".
Example: $WIMTA,024,C*33
5.1 Possible settings for command NO
parameter value
unit of velocity
talker
address
MWV
output
MTA
output
0 … 7
disabled
disabled
8
knots
$II
enabled
disabled
9
mph
$II
enabled
disabled
10
km/h
$II
enabled
disabled
11
m/s
$II
enabled
disabled
12
knots
$WI
enabled
disabled
13
mph
$WI
enabled
disabled
14
km/h
$WI
enabled
disabled
15
m/s
$WI
enabled
disabled
16
knots
$II
disabled
enabled
17
mph
$II
disabled
enabled
18
km/h
$II
disabled
enabled
19
m/s
$II
disabled
enabled
20
knots
$WI
disabled
enabled
21
mph
$WI
disabled
enabled
22
km/h
$WI
disabled
enabled
23
m/s
$WI
disabled
enabled
24
knots
$II
enabled
enabled
25
mph
$II
enabled
enabled
26
km/h
$II
enabled
enabled
27
m/s
$II
enabled
enabled
28
knots
$WI
enabled
enabled
29
mph
$WI
enabled
enabled
30
km/h
$WI
enabled
enabled
31
m/s
$WI
enabled
enabled