MD 9225 Clamp Meter Series Specifications
14
5 Specifications
Display: 3-3/4 digits 4000 counts LCD display(s)
Update Rate: 3 per second nominal
Polarity: Automatic
Low Battery: Below approx. 2.4V
Operating Temperature: 0
o
C to 40
o
C
Relative Humidity: Maximum relative humidity 80% for temperature up
to 31
o
C decreasing linearly to 50% relative
humidity at 40
o
C
Altitude: Operating below 2000m
Storage Temperature: -20
o
C to 60
o
C, < 80% R.H. (with battery removed)
Temperature Coefficient: nominal 0.15 x (specified accuracy)/
o
C @(0
o
C -
18
o
C or 28
o
C -40
o
C), or otherwise specified
Sensing : True RMS
Safety : Meets IEC/UL/EN61010-1 Ed. 3.0, IEC/EN61010-
2-032 Ed. 3.0, IEC/EN61010-2-033 Ed. 1.0,
IEC/UL/EN61010-031 Ed. 1.1 & CAN/CSA-C22.2
No. 61010-1-12 Ed. 3.0:
Category: III 600 Volts ac & dc.
Transient protection: 6.5kV (1.2/50s surge)
Pollution degree: 2
E.M.C.: Meets EN61326-1:2006 (EN55022, EN61000-3-2,
EN61000-3-3, EN61000-4-2, EN61000-4-3,
EN61000-4-4, , EN61000-4-5, EN61000-4-6,
EN61000-4-8, EN61000-4-11)
In an RF field of 3V/m:
Capacitance function is not specified
Other function ranges: Total Accuracy = Specified Accuracy + 50igits
Performance above 3V/m is not specified
Overload Protections:
Clamp-on jaws: DC/AC 400A rms continuous
+ & COM terminals (all functions): 600 VDC/VAC rms
Power Supply: standard 1.5V AAA Size (NEDA 24G or IEC R03)
battery X 2; or 1.5V AAA Size (NEDA 24A or IEC
LR03) alkaline battery X 2
Power Consumption: 11mA typical for ACA/DCA and 2.9mA typical for
other functions
APO Timing: Idle for 30 minutes
APO Consumption: typical 190A
Dimension: L188mm X W63mm X H40mm
Weight: 218 gm approx
Jaw opening & Conductor diameter: 30mm max
Accessories: Test lead pair, batteries installed, user's manual,
soft carrying pouch, & banana plug type-K
thermocouple
Optional accessories: AMD 9024 banana plug to type-K socket plug
adaptor
Electrical Specifications