Functional description
100.000-count Graphical Multimeters 25
Functional description (cont’d)
MATH function
MTX 3282
MTX 3283
The MATH function (y = Ax + B) enables measuring any physical quantity in:
- Volts (0 - 10 V process or high-voltage probe, for example)
- Amperes (current loop 4 - 20 mA or current clamp, for example)
- Frequency (output measurement, rotation speeds, for example)
- Ohms (resistive position, for example)
and converting it and to assign the adequate unit, to obtain the direct reading of
the original parameter on the instrument. Depending on the parameter measured,
the device calculates the related MATH function.
The programming takes place in 4 phases:
1.
Selection of parameter X measured (V, A, , Hz)
2.
Definition of coefficient A of function y = Ax + B
3.
Definition of coefficient B of function y = Ax + B
4.
Definition of physical unit to be displayed
Coefficients A, B and the unit are programmable for each amount
measured (V, A,
, Hz).
Adjustment of
MATH function
The MATH mode is configured in the MATH function in the Configuration Menu.
Opening of Configuration Menu using key .
Selection of function MATH Func. using keys
.
Validation of MATH Func. using key
.
Selection of
function to be
adjusted
Selection of Function menu using keys
.
Selection of measurement (V, A, , Hz) using key
.
Default function V.
u
General
Measure
Func.
MATH function
settings
Func. MATH
Configuration
MATH function
Coef A
Coef B
Function to be set
Unit
Function V