EasyManua.ls Logo

METRON QA-45 - QA-45 Control and Calibration; QA-45 Control and Calibration Test Equipment; QA-45 Control and Calibration Preparation; QA-45 Control and Calibration References

METRON QA-45
74 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
6. Control and Calibration
This chapter explains the QA-45 maintenance procedures, including
testing and calibration.
6.1 Required Test
Equipment
Digital multimeter, 10 uV resolution, 0.1% accuracy.
Frequency counter
Oscilloscope
Variable VIA power supply
10 V (+0.01 v) power source
Pulse generator: square pulse, 10 ms width, 10V amplitude, 80
pulses per. minutes (for pacer module).
6.2 Preparation
Set the switches on QA-45 to the following positions:
Mode: Low
Power: Off
Connect a power supply to the battery eliminator input on QA-45.
Adjust the power supply to 9V (±0.2v) with a power limitation of
200 mA (±50 mA).
6.3 References
The function keys are numbered 1 to 5, with switch 1 farthest to left.
6.4 Test
1. Set power to ON. Wait 3 seconds, and measure the current from
the power supply. Requirement: 68 mA (±5 mA).
2. Adjust P2 on the processor board to obtain the best possible con-
trast on the display. The display should show the main menu
and the result with 0-data. Press function switch 5 and check
that QA-45 changes between various menus. Adjust P103 out of
limit.
3. Set power to OFF, and then back to ON after about 1 second.
The display should show the software version number for a
brief period, before showing the main active display. (By test
under production a additional item shall be used.)
4. Measure the operating voltages in QA-45 with the multimeter.
The following values are acceptable:
Test point
- +
Level Maximum
Deviation
1

Related product manuals