When the defibrillator is charged, discharge it through the instru-
ment.
5. Charging time will be shown in the display as ‘Chrg T: xx.x MS’
under RESULT.
4.6 Shock Advisory
Algorithm Test
1. This tests the analysis and prompting of automatic and semi-au-
tomatic defibrillators. A series of arrhythmia is available for
analysis by the defibrillator that should then prompt the user to
‘shock’ of ‘no shock,’ in accordance with national and interna-
tional guidelines, as shown below:
ASYS No shock
SVTa_90 No shock
PVT_140 No shock
MVT_140 No shock
CVF Shock
FVF Shock
PVT_160 Shock
MVT_160 Shock
2. Select ADV. ALG. (F2) from the main menu.
3. The Advisory Algorithms Menu opens.