Neutral electrode test setup in compliance with IEC 601.2.2,
sec. 19.101b, fig, 104 and sec. 19.102. (Adopted by
ANSI/AAMI HF18-1993)
4.3 REM Alarm
This test ensures that the ESU will sound an alarm if the resistance
between the two neutral electrodes exceeds your specified limit. The
program directs the OA-ES to gradually increase the resistance. At a
certain value, the ESU should sound an alarm. Test setup for ESU
REM alarm test.