Operation
optoNCDT ILR 102x/110x/115x
Analog-Setup
Q
A
Q
B
S T
Setting analog output
Press the Toggle key until the menu
LEDs SA1 and SB1 light up.
Position the reflector at the desired 0 %
point (4 mA), the press the Set key.
Q
A
Q
B
S T
Press the Toggle key until the menu
LEDs SA2 and SB2 light up.
Position the reflector at the desired
100 % point (20 mA), then press the
Set key.
If the set distance between the 0 % and the
100 % point is less than 600 mm the sensor
sets the analog output automatically to
600 mm. The middle of the measuring range
is placed automatically between the two set
positions.
Q
A
Q
B
S T
Switching between Fast and Slow mode
Press the Toggle key until the LED
SLOW lights up.
Press Set key than the sensor switch-
es between SLOW and FAST mode.
LED SLOW flashes slow: SLOW mode
LED SLOW flashes fast: FAST mode
Q
B
-Setup
Q
A
Q
B
S T
Setting switching point Q
B
Press the Toggle key until the menu
LED SB1 lights up.
Position the reflector at the desired
switching point, then press the SET
key.
Q
A
Q
B
S T
Setting switching window Q
B
Press the Toggle key until the menu
LED SB1 lights up.
Position the reflector at the desired
first switching point, the press the Set
key.
Q
A
Q
B
S T
Press the Toggle key until the menu
SB2 lights up.
Move the reflector to the desired
second switching point, then press the
Set key.
Q
A
Q
B
S T
Invert switching output Q
B
Press the Toggle key until the menu
LEDs SB1 and SB2 light up.
Press the Set key.
The output signal is inverted.