Maintenance VacPrep 061
3-8 Feb 2013
2. If you are cleaning the fittings, disassemble them and clean the components with isopropyl
alcohol (IPA). If the fitting O-ring is cracked or broken, replace the O-ring.
3. Remove the two screws that hold the port nipple to the degasser. Remove the port nipple,
dust frit and O-ring. Clean all components with IPA (an ultrasonic bath may be used). If
the dust filter O-ring is cracked or broken, replace it. Reattach the dust frit, O-ring, and
port nipple.
4. Attach the cleaned or replacement fittings you removed in step 1 by pressing the flexible
tubing completely over the barbed nipple on the end of the fitting. Loosely attach the
knurled retaining nut, ferrule, and O-ring.
Ensure adequate ventilation when using solvents for cleaning purposes.
Dust Frit
Dust Frit O-ring
Port Nipple