18 Xpress Basics
Observe the following carefully during quick test
• Arm centers on spindle
• Disc is dropped in center of drive writer tray
• Disc is dropped onto output spindle
If Printer is installed
• Disc is dropped in center of printer tray
• Disc is dropped directly onto output spindle
If the single initial test fails, run the continuous test to locate the
problem—Make sure there are several discs on spindle 3. Click the
continuous test button at the lower right to run the same test
continuously.
Disc handling problems are not always repeatable. Accurate calibration
is important to achieve reliable operation. It is recommended that you
run the continuous test through several hundred discs to insure proper
calibration.
After starting the test, this button name will automatically change to
Stop Continuous. To stop the test, click this button.
If the problem recurs, calibration (adjustment) of the system is necessary.
(See the following chapter, Calibrating Xpress for complete
instructions.)