ScanMaker 9800XL Plus User Guide 33
Calibration Setup
With the target properly positioned on the scanner, turn on your scanner and
let it warm up for about five minutes. Launch the MSP program; the MSP main
window appears.
A. Choose the scanner model that you are calibrating.
B. Choose the correct target media. If you are calibrating the reflective target, select
Reflective. Otherwise, select Positive.
C. Select the date code and target type (or reference
number) from the drop-down list that matches your
target. You can verify this information by looking at the
left corner of the target’s bottom margin.
If the desired data file is not present in the drop-down menu, install the IT8
Calibration Data to update the target profiles in your system.
D. When all the settings are done, click the Start (or Start Profiling) button. The
calibration window will appear, and an initial preview of the target is performed.
C
D
A
B