Maintenance and care
This machine should be inspected in
accordance with local and national
safety regulations after every 1000
operating hours, or every 6 months by a
Miele approved service technician.
A service will cover the following:
Electrical safety tests
Door mechanism and door seal
Any screw connections and
connectors in the wash cabinet
Water inlet and drainage
Internal and external dispensing
systems
Spray arms
Filter combination
Sump including drain pump and
non-return valve
Steam condenser
All mobile units, baskets and inserts
The drying unit
and where applicable:
Any printer connected to the
machine.
The following operational tests will be
carried out within the framework of the
maintenance:
A programme will be run as a test
run
Thermo electrical measurements will
be taken
Seals will be tested for water
tightness
All relevant measuring systems will
be safety tested including error
message displays.
Safety devices.
Routine checks
Before the start of each working day,
the user must carry out a number of
routine checks.
The following need to be inspected:
All filters in the wash cabinet
The spray arms in the machine and
in any mobile units or baskets
The wash cabinet and the door seal
Mobile units, baskets and inserts.